510 Error codes
Page: 64
SCION Instruments
436-GC/456-GC Service Manual Revision B February 2019
Code Error Description Error level Cause Condition(s) GC action Action needed
device. 230 V GC is using 110 V heaters
H2 flow set to 0.0 mL/min
110 Zone 1 too slow response Error
Sensor not used (wrong used), mismatch heated
device. 230 V GC is using 110 V heaters
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (1)
111 Zone 2 too slow response Error
Sensor not used (wrong used), mismatch heated
device. 230 V GC is using 110 V heaters
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (2)
112 Zone 3 too slow response Error
Sensor not used (wrong used), mismatch heated
device. 230 V GC is using 110 V heaters
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (3)
113 Zone 4 too slow response Error
Sensor not used (wrong used), mismatch heated
device. 230 V GC is using 110 V heaters
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (4)
114 Zone 5 too slow response Error
Sensor not used (wrong used), mismatch heated
device. 230 V GC is using 110 V heaters
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (5)
115 Zone 6 too slow response Error
Sensor not used (wrong used), mismatch heated
device. 230 V GC is using 110 V heaters
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (6)
Too Fast Response (zone)
116 Column Oven too fast response Error
Sensor not used (wrong used), mismatch heated
device. 110 V GC is using 230 V heaters
Electronics off (all)
H2 flow set to 0.0 mL/min
Check Column Oven zone
117 Zone 1 too fast response Error
Sensor not used (wrong used), mismatch heated
device. 110 V GC is using 230 V heaters
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (1)
118 Zone 2 too fast response Error
Sensor not used (wrong used), mismatch heated
device. 110 V GC is using 230 V heaters
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (2)
119 Zone 3 too fast response Error
Sensor not used (wrong used), mismatch heated
device. 110 V GC is using 230 V heaters
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (3)
120 Zone 4 too fast response Error
Sensor not used (wrong used), mismatch heated
device. 110 V GC is using 230 V heaters
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (4)
121 Zone 5 too fast response Error
Sensor not used (wrong used), mismatch heated
device. 110 V GC is using 230 V heaters
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (5)
122 Zone 6 too fast response Error
Sensor not used (wrong used), mismatch heated
device. 110 V GC is using 230 V heaters
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (6)
Fan or Heater
123
No Fan or Heater detected
Error
Hardware error
-
-
124
Watch Dog Heater activated
Error
Hardware error
-
-
EFC Hardware Errors
125
Front EFC Flow Unit Hardware Error
126
Middle EFC Flow Unit Hardware Error
127
Rear EFC Flow Unit Hardware Error
128
Front EFC Pressure Unit Hardware Error
129
Middle EFC Pressure Unit Hardware Error
130
Rear EFC Pressure Unit Hardware Error
131
Front Detector EFC Channel 1 Hardware Error
132
Front Detector EFC Channel 2 Hardware Error
133
Front Detector EFC Channel 3 Hardware Error
134
Middle Detector EFC Channel 1 Hardware Error
135
Middle Detector EFC Channel 2 Hardware Error
136
Middle Detector EFC Channel 3 Hardware Error
137
Rear Detector EFC Channel 1 Hardware Error
138
Rear Detector EFC Channel 2 Hardware Error
139
Rear Detector EFC Channel 3 Hardware Error
140
Aux 1 EFC Hardware Error
141
Aux 2 EFC Hardware Error

Table of Contents