436-GC/456-GC
510 Error codes
Page: 68
SCION Instruments
436-GC/456-GC Service Manual Revision B February 2019
Code Error Description Error level Cause Condition(s) GC action Action needed
428 Sampler Tower stuck on sensor Alert (errors not reproducible)
Electronics off (all)
H2 flow set to 0.0 mL/min
-
429 Sampler Carrousel stuck on sensor Alert (errors not reproducible)
Electronics off (all)
H2 flow set to 0.0 mL/min
-
434
Sampler EDS Checksum is bad
Alert
-
-
-
436
IOP Communication failure
Error
IOP Communication failure
Notification Message?
Replace/fix IOP
437
Battery voltage out of range
Event
Battery voltage < 2V or > 4V error
-
-
439
PGA Calib Failed
Error
IOPSH (measuring system)
-
Recalibrate PGA
440
Temp Calib Failed
Error
-
-
Recalibrate temp
441
Temperature Calibration Failed
442
Press out of tolerance
Error
Ambient pressure < 15kPa or > 115pKa error
-
-
443
Boardtemp out of tolerance
Error
Currently in use, only logged
-
GC main board
Out of Tolerance (red_id)
444
P10V ref out of tolerance
Error
444 to 465 Refvolt out of Tolerance (all voltages on
mainboard, part of measuring system) Hazardous by
default, basis of all measurement in system.
-
Check power supply requirements/
replace main board electronics
445
AIN10V out of tolerance
Error
-
446
P5V ref out of tolerance
Error
-
447
P2V5 ref out of tolerance
Error
-
448
P24V Prim out of tolerance
Error
-
449
P24V Sec out of tolerance
Error
-
450
N24V out of tolerance
Error
-
451
P5V D out of tolerance
Error
-
452
P5V out of tolerance
Error
-
453
N5V out of tolerance
Error
-
454
P15V out of tolerance
Error
-
455
N15V out of tolerance
Error
-
456
P5V DAC out of tolerance
Error
-
457
VREF1 out of tolerance
Error
-
458
VREF0 out of tolerance
Error
-
Out of Tolerance (cts_id)
459 CTS1 out of tolerance Error -
Electronics off (all)
H2 flow set to 0.0 mL/min
-
460
CTS2 out of tolerance
Error
-
-
461
CTS3 out of tolerance
Error
-
-
462
CTS4 out of tolerance
Error
-
-
463
CTS5 out of tolerance
Error
-
-
464
CTS6 out of tolerance
Error
-
-
465
Tref out of tolerance
Error
-
-
466
Step motor moving
Error
-
-
Temperature
470
Maximum Setup Temperature Exceeded
471
Maximum Temperature Exceeded
473
Minimum Temperature Exceeded
474
Open Sensor
476
Shorted Sensor
477
Thermal Runaway
478
Too Fast Response
479
Too Slow Response