__________________ Vortex M6
12
NOTE:Improperinstallationorflexingofweakfloorwillcauseexcessivevibration.Donot
slidethe Vortex M6 across flooronce the leveling legs have beenextended,aslegsand
basecouldbecomedamaged.
3.4.5.9 Verifythattheinstrumentdoesnotrock.
3.4.6 PlugintheInstrument
ElectricalRequirements
220Volt±10%,50/60HertzwithEarth/GroundPlug
WhenpluggingintheVortexM6:
•DONOToverloadcircuits.
•DONOTuseanextensioncord.
•DONOTuseanadapter.
•DONOToperateotherappliancesonthesamecircuit.Useseparately fused 15 Amp
circuits.
The Vortex M6 is designed to be operated on a separate branch, polarized, three‐wire,
effectiveearth/ground,220Volt,50/60Hertz,AC(alternatingcurrent),circuitprotectedbya
15amperefuse,equivalentfuseorcircuitbreaker.
Thepowercordshouldbepluggeddirectlyintoaneffectiveearth/groundreceptaclerated
220VoltsAC(alternatingcurrent)15Amps.