Troubleshooting seca
03.04.2013/JEG 16 (24) 30-34-00-864
3.2 Scale error messages
The scales have an error message feature which indicates possible malfunctions and
defects. These error messages are described below.
3.3 Error messages
[334|335|382|383|384|385|833|834|835|874|875|876|877|878|952|956]
If a malfunction or defect occurs during operation of the scale, it is shown on the display
as an error message. Error messages are displayed in the following form:
"E xx"
E: Error
xx: Error number (see Table 4)
Displayed Cause Remedy
E 11
Load cell defective, supply
leads damaged or electronics
defective
Diagnosis with seca serva:
Check load cell connections
Poss. replace load cell or
electronics
Check preload on scale
E 12
Preload too great, load cells
damaged or electronics
defective
Diagnosis with seca serva:
Poss. replace load cell or
electronics
E 13
Limit for automatic zero tracking
reached
Place load of >0.5d/s on scale
E 16
Error when determining zero
point on switch-on
Protect scale from vibration /
shocks during start-up