Seca 334 LC - Error Messages [Models 769-899]

Seca 334 LC
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting seca
03.04.2013/JEG 17 (24) 30-34-00-864
seca serva: Recalibrate scale
E 40 EEPROM data incorrect
Replace electronics
Table 4: Error messages [334|335|382|383|384|385|833|834|835|874|875|876|877|878|952|956]
3.4 Error messages [769|799|780|869|899]
If a malfunction or defect occurs during operation of the scale, it is shown on the display
as an error message. Error messages are displayed in the following form:
"E mm.xx"
E: Error
mm: Number of the module reporting the error
(10 = weight calculation, 11 = electronic display)
xx: Error number (Table 5)
Displayed Cause Remedy
E XX. 11
Load cell defective, supply
leads damaged or electronics
defective
Diagnosis with seca serva:
Check load cell connections
Poss. replace load cell or
electronics
Check preload on scale
E XX.12
Preload too great, load cells
damaged or electronics
defective
Diagnosis with seca serva:
Poss. replace load cell or
electronics
E XX.13
Limit for automatic zero tracking
reached
Place load of >0.5d/s on scale