66 •
2. Press the switch menu button.
3. Press the
settings button.
The
User dialog window appears.
The verification counter reading (in this case: 1) is displayed.
4. Compare the content of the verification counter output with the number
given on the verification counter sticker.
Both numbers have to match for the verification to be valid. If the sticker and
the verification counter do not match, the scale must be subsequently verified.
Please contact your service partner or seca service.
Once the scale has been subsequently verified, a new, updated verification
counter sticker is used to identify the verification counter reader. The person
authorized to perform the subsequent verification secures this sticker with an
additional seal. The verification counter sticker can be ordered from seca
service.
11.MAINTENANCE (UNVERIFIED MODEL)
On leaving the factory, your seca device has an accuracy of ±0.15% or bet-
ter. To preserve this level of accuracy, the product must be set up with care
and maintained on a regular basis.
The device’s measurement technology for bioimpedance analysis (BIA) must
be checked every two years. We recommend performing maintenance of the
whole device as part of this check.
NOTICE!
Incorrect measurements as a result of poor servicing
Please only have servicing and repairs performed by an authorized
service partner.
You can find service partners in your area at www.seca.com or by
sending an e-mail to service@seca.com.