EasyManua.ls Logo

Seca 6641321108 - Radio Error Message [374|376|378|703|704|284|285|634|635|644|645|656|657|664|665| 674|675|676|677|684|685|954|957|959|963]

Seca 6641321108
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Errors seca
14.11.12/RLEU/ARI/MRE 26/40 30-34-00-812c
Er:X:43 EEPROM defective Replace the
electronics
Er:X:45 EEPROM defective Replace the
electronics
Er:X:57 Quartz does not oscillate Replace the
electronics
3.6 Radio Error Message
[374|376|378|703|704|284|285|634|635|644|645|656|657|664|665|
674|675|676|677|684|685|954|957|959|963]
Display: Er:x:nr Electronics detecting a fault.
X : Module number
No Fault number
Scale displays Cause Remedy
Er:X:71
Radio module is
switched off
Switch on the radio
module via the
menu (see
operating
instructions)
Er:X:72
Error in teach-in mode Repeat teach-in for
the devices (see
operating
instructions)

Table of Contents