Why does "StOP" appear on my Seca 634 display?
- MMichael Smith DDSSep 12, 2025
The "StOP" message appears because the maximum load for the scale has been exceeded. Remove some of the load from the scale.
Why does "StOP" appear on my Seca 634 display?
The "StOP" message appears because the maximum load for the scale has been exceeded. Remove some of the load from the scale.
What does it mean when the display "bAtt" appears on Seca 634 Scales?
The "bAtt" message on the display indicates that the batteries are flat. You should insert new batteries.
How to fix "Er:X:11" error on Seca Scales?
The "Er:X:11" error indicates that the scale has too high a load, or the load is unevenly distributed. To resolve this, take some of the load off the scale or distribute the weight more evenly. If that doesn't work, restart the scale.
What does "Er:X:12" mean on Seca 634 Scales and how to resolve it?
The "Er:X:12" error appears when the scale was switched on with too high a load. Remove the load from the scale. If the issue persists, restart the scale.
What to do if Seca Scales displays "tEMP"?
The display "tEMP" indicates that the ambient temperature is too high or too low. Set up the scale in an environment where the temperature is between +10 °C and +40 °C. You can also wait approximately 15 minutes for the device to adjust to the ambient temperature.
How to fix Seca Scales when no weight is displayed with a load?
If no weight is displayed when there is a load on the scale, check whether the scale is switched on and whether batteries are inserted.
How to resolve "Er:X:16" error on Seca Scales?
The "Er:X:16" error means that the scale was caused to oscillate and the zero point could not be determined. Restart the scale.
Why does my Seca Scales not show 0.00 before weighing?
If 0.00 does not appear before weighing, it could be that a load was already on the scale when it was switched on. Remove the load from the scale. If this doesn't work, switch the scale off and back on again.
What to do when Seca 634 transmits measurement results and beeps after the first switch on?
This happens because the device was unable to send measurement results to the wireless receiver. Ensure that the scale is integrated into the wireless network. Also, keep HF equipment at least 1 meter away from the transmitters and receivers, as reception may be interfered with by HF equipment.
Why do I see only the "SYS" item in the "rF" menu of my Seca 634?
If you only see the "SYS" item in the "rF" menu, the wireless module is deactivated. Activate the wireless module.
Graduation | 100 g |
---|---|
Display | LCD |
Automatic switch-off | Yes |
Type | Digital scale |
Capacity | 300 kg |
Functions | BMI calculation |
Power supply | Power adapter |
Platform size | 320 x 320 mm |
Weight units | kg, lb |
Details how DMS modules use load cells and ADCs for weight measurement.
Lists possible causes and remedies for the scale failing to start.
Identifies causes and solutions for inaccurate weight readings.
Provides troubleshooting steps for segment test issues or fluctuating values.
Lists potential issues and fixes when the scale doesn't respond to buttons.
Addresses errors like 'Er:X:10', 'Er:X:11', 'Er:X:12' during weighing.
Details faults like kg value calculation, battery, temperature, and bus faults.
Step-by-step guide on switching to calibration mode and performing recalibration.
Provides a user action and result summary for the recalibration process.
Details the equipment and steps for PC-based diagnosis or recalibration.
Details the process of dismantling the scale platform, frame, and electronics housing.