Function check • 85
3. Disinfect the device by moistening a soft cloth in
disinfectant and wiping the device over with it.
:
7.3 Sterilizing
This device may not be sterilised.
8. FUNCTION CHECK
► Perform a function check before each use.
A complete function check includes:
• a visual inspection for mechanical damage
• a visual inspection and function check of the display
• function check of all the controls shown in the
section entitled "Overview".
• a function check of the optional accessories
If you find faults or deviations during the function
check, as a first step try to remedy the fault with the aid
of the section entitled "Troubleshooting" from page 86.
CAUTION!
Bodily injury
If you find faults or deviations during the func-
tion check which you are unable to remedy with
the aid of the section entitled "Troubleshooting"
from page 86, you may not use the device.
► Have the device repaired by seca service or
an authorized specialist dealer.
Interval Component
Prior to every measurement
with direct skin contact
Weighing platform
After every measurement
with direct skin contact
Weighing platform
If required
Display
Membrane keyboard