Service Manual Replacement instructions
02.05.05 / Rei 2/4 30-34-00-720
2.4 Replacing display board "AP", membrane keyboard "FT" and battery connection "BA"
(page 4)
Unscrew the 4 M6 countersunk screws "C1" from the side covers.
Unsolder the cable connections from the parts to be replaced.
Display board "AP": Slacken the four fixing screws "P2" and detach the plug "S3" of the cable
connection from the bottom part.
Membrane keyboard "FT": Carefully detach the membrane keyboard "FT" (glued on) and pull the
panel housing "BG" off to the left.
3. Closing the scale
3.1 Reassemble the scale in reverse order, following the instructions given in sections 2.4 to 1.1.
Warning !
When the DMS Electronic module and the platform load cell are replaced, the scale must be re-
adjusted. Follow instructions no. 30-34-00-603.
When replacing electronic components, observe cabling diagram 08-02-06-032.
3.2 Apply for official calibration of the scale.