Errors seca
14.11.12/RLEU/ARI/MRE 22/40 30-34-00-812c
3.2 Error Symptoms for Head Slider [264|274|284|285]
Error description
Possible causes Remedy
Battery not inserted or flat Check the batteries
Voltage supply defective See section 4.1
Head slider does
not start
Key panel defective or not
connected
Check cable
connections, check
cable and key panel
for interruptions,
replace if necessary
Insert scale defective /
missing / scale top and
bottom reversed
Check the insert scale
Head slider not adjusted
correctly
See section 2.2
Light barriers soiled
Clean the head slider
barriers with a cloth
Key unit not engaged
correctly
Check the key unit
Head slider does
not display a
length or an
incorrect length