Replacement
seca 674, 675
08.02.2013/MRE Seite 7 (9) 30-34-00-859_E1
1.3 Setting the overload protection devices
To protect the scale from overload, four overload protection devices are provided.
Set the specified values for the relevant model in accordance with the table (Table 1);
use gauges for this purpose.
The settings must be very accurate to prevent a forced shunt.
Model Dimension x =
634/635/644/645 7 mm
664/665 3.5 mm
674/675 7 mm
676/677 3.5 mm
684/685 7 mm
Table 1: Dimensions to be set for the overload protection device for multi-function scales
Figure 4: Setting the overload protection devices