Function check • 13
English
NOTICE!
Damage to device
Inappropriate detergents and disinfectants may damage the sensitive
surfaces of the device.
► Use only disinfectants free of chlorine and alcohol which are
explicitly suitable for acrylic sheet and other sensitive surfaces
(active ingredient: quaternary ammonium compounds, for example).
► Do not use caustic or abrasive detergents.
► Do not use organic solvents (e.g. white spirit or petroleum spirit).
6.1 Cleaning
► Use a soft cloth dampened with mild soapsuds to clean the surfaces of
the device.
6.2 Disinfecting
1. Check that your disinfectant is suitable for sensitive surfaces and acrylic
sheet.
2. Follow the instructions on the disinfectant.
3. Disinfect the device by moistening a soft cloth in disinfectant and wiping
the device over with it.
:
6.3 Sterilizing
This device may not be sterilised.
7. FUNCTION CHECK
► Perform a function check prior to each use.
A complete function check includes:
• visual inspection for mechanical damage
• checking the alignment of the device
• visual and function check of the display elements
• function check of all the controls shown in the section entitled “Overview”
• function check of optional accessories
If you notice any faults or deviations during the function check, first try to
resolve the error with the aid of the section entitled “What do I do if ...” in this
document.
Interval Component
Prior to every
measurement with direct
skin contact
Weighing platform
After every measurement
with direct skin contact
Weighing platform
If required Display