14 •
CAUTION!
Personal injury
If you notice any faults or deviations during the function check which
cannot be resolved with the aid of the section entitled “What do I do
if ...” in this document, you may not use the device.
► Have the device repaired by seca Service or by an authorized
service partner.
► Follow the section entitled “Servicing” in this document.
8. WHAT DO I DO IF...?
9. MAINTENANCE
On leaving the factory, your seca scale has an accuracy of ±0.15 %/± 50 g/±
0.22 lbs. To preserve this level of accuracy, the product must be set up with
care and maintained on a regular basis. Depending on how frequently the
scale is used, we recommended servicing at intervals of 3 to 5 years.
CAUTION!
Incorrect measurements as a result of poor servicing
►
Please only have servicing and repairs performed by an authorized
service partner.
► You can find service partners in your area at www.seca.com or by
sending an e-mail to service@seca.com.
Problem Cause/solution
... no weight is displayed when there
is a load on the scale?
The scale has no power supply.
- Check whether the scale is switched on
- Check whether batteries are inserted
... the scale switches on, during
transport for example?
The start key has been pressed. As the scale has a very low energy
consumption, this will have very little effect on battery life.
- Set the On/off switch to "OFF" for relatively long transport distances
... a weight value appears in the
display of the scale following
transport or a change of batteries,
even without a load on it?
The scale has determined an incorrect zero point.
- Start the scale again. The scale will then work normally again
...
does not appear before
weighing?
A load was already on the scale before it was switched on.
- Remove the load from the scale
- Switch the scale off and back on again
... - - - - appears before weighing
instead of
?
A load was already on the scale before it was switched on.
- Remove the load from the scale
- Switch the scale off and back on again
... one segment is illuminated either
continuously or not at all?
The corresponding point has a fault.
- Inform the maintenance service
… the display appears?
Battery voltage is dropping.
- Insert new batteries as soon as possible
… the display
appears?
The batteries have discharged.
- Insert new batteries
… the display
appears?
Maximum load has been exceeded.
- Remove the load from the scale
... the display
and a number
appear?
The scale has been incorrectly loaded.
- Re-start the scale after it has switched off automatically. The scale will
then work normally again.
- If this is not the case, interrupt the power supply by switching the scale
off briefly at the On/off switch. If this measure is also unsuccessful, inform
the maintenance service.