CAUTION!
Personal injury
If you notice any faults or deviations during the function check which
cannot be resolved with the aid of the section entitled “What do I do
if ...” in this document, you may not use the device.
Have the device repaired by seca Service or by an authorized
service partner.
Follow the section entitled “Servicing” in this document.
8. WHAT DO I DO IF...?
9. MAINTENANCE
On leaving the factory, your seca scale has an accuracy of ±0.15 %/± 50 g/±
0.22 lbs. To preserve this level of accuracy, the product must be set up with
care and maintained on a regular basis. Depending on how frequently the
scale is used, we recommended servicing at intervals of 3 to 5 years.
CAUTION!
Incorrect measurements as a result of poor servicing
Please only have servicing and repairs performed by an authorized
service partner.
You can find service partners in your area at www.seca.com or by
sending an e-mail to service@seca.com.
Problem Cause/solution
... no weight is displayed when there
is a load on the scale?
The scale has no power supply.
- Check whether the scale is switched on
- Check whether batteries are inserted
... the scale switches on, during
transport for example?
The start key has been pressed. As the scale has a very low energy
consumption, this will have very little effect on battery life.
- Set the On/off switch to "OFF" for relatively long transport distances
... a weight value appears in the
display of the scale following
transport or a change of batteries,
even without a load on it?
The scale has determined an incorrect zero point.
- Start the scale again. The scale will then work normally again
...
 does not appear before
weighing?
A load was already on the scale before it was switched on.
- Remove the load from the scale
- Switch the scale off and back on again
... - - - - appears before weighing
instead of
?
A load was already on the scale before it was switched on.
- Remove the load from the scale
- Switch the scale off and back on again
... one segment is illuminated either
continuously or not at all?
The corresponding point has a fault.
- Inform the maintenance service
… the display appears?
Battery voltage is dropping.
- Insert new batteries as soon as possible
… the display
 appears?
The batteries have discharged.
- Insert new batteries
… the display
 appears?
Maximum load has been exceeded.
- Remove the load from the scale
... the display
 and a number
appear?
The scale has been incorrectly loaded.
- Re-start the scale after it has switched off automatically. The scale will
then work normally again.
- If this is not the case, interrupt the power supply by switching the scale
off briefly at the On/off switch. If this measure is also unsuccessful, inform
the maintenance service.