10.TECHNICAL DATA
10.1 General technical data
10.2 Weighing data
General technical data
Dimensions
Depth
Width
Height
360 mm
321 mm
60 mm
Net weight approx. 4.0 kg (8.8 lbs)
Temperature range
Operation
Storage
Transport
+10 °C to +40 °C (50 °F to 104 °F)
-10 °C to +65 °C (14 °F to 149 °F)
-10 °C to +65 °C (14 °F to 149 °F)
Air pressure
Operation
Storage
Transport
700 – 1060 hPa
700 – 1060 hPa
700 – 1060 hPa
Humidity
Operation
Storage
Transport
30 % – 80 %, no condensation
0 % – 95 %, no condensation
0 % – 95 %, no condensation
Height of digits 20 mm
Power supply
Power supply unit
- Supply voltage
- Insulated device, protection class II
(EN 60601-1)
Batteries
- Supply voltage
12 V
9 V
Maximum current consumption typically 50 mA
Medical device in accordance with
Directive 93/42/EEC
Class I with measuring function
EN 60601-1:
electrical medical device, type B
Type of protection IP20
Operating mode Continuous duty
Weighing technology data for seca 874
Maximum load 200 kg (440 lbs)
Accuracy
0 to 50 kg (0 lbs to 110 lbs)
50 kg to 200 kg (0 lbs to 440 lbs)
± 100 g (0.22 lbs)
± 0.15 %
Graduations
0 to 150 kg (0 lbs to 330 lbs)
150 kg to 200 kg (330 lbs to 440 lbs)
50 g (0.1 lbs)
100 g (0.2 lbs)