12 •
4. To deactivate the HOLD function, press the hold key
The symbol and the "HOLD" message are no longer displayed.
Switching off the scale
Press the start key.
NOTE:
The scale switches off automatically after a short time if there is no
load on it.
6. HYGIENE TREATMENT
WARNING!
Electric shock
The device is not de-energized when the on/off key is pressed and the
display goes out. Use of fluids on the device may cause electric
shock.
Ensure that the device is switched off before performing any hygiene
treatment.
Disconnect the power supply connector before performing any
hygiene treatment.
Ensure that no fluids penetrate the device.
CAUTION!
Damage to device
Inappropriate detergents and disinfectants may damage the sensitive
surfaces of the device.
Use only disinfectants free of chlorine and alcohol which are
explicitly suitable for acrylic sheet and other sensitive surfaces
(active ingredient: quaternary ammonium compounds, for example).
Do not use white spirit or petroleum spirit.
Only use disinfectants suitable for sensitive surfaces. Suitable
disinfectants are available from specialist dealers.
6.1 Cleaning
Use a soft cloth dampened with mild soapsuds to clean the surfaces of
the device.
6.2 Disinfecting
1. Check that your disinfectant is suitable for sensitive surfaces and acrylic
sheet.
2. Follow the instructions on the disinfectant.
3. Disinfect the device by moistening a soft cloth in disinfectant and wiping
the device over with it.
:
kg
6.50
start key
Interval Component
Prior to every
measurement with direct
skin contact
Weighing platform