Safety precautions • 5
English
Preventing injuries and infections WARNING!
Injury from falls
Ensure that the device is positioned firmly and level.
Route connecting cables (if present) in such a way that neither user
nor patient can trip over them.
The device is not designed as a standing aid. Assist people with
limited motor skills when they are getting up, e.g. from a wheelchair.
Make sure that the patient does not step onto and off the weighing
platform right at the edges.
Make sure that the patient steps onto and off the weighing platform
slowly and safely.
WARNING!
Danger of slipping
Ensure that the weighing platform is dry before the patient steps
onto it.
Ensure that the patients feet are dry before he or she steps onto the
weighing platform.
Make sure that the patient steps onto and off the weighing platform
slowly and safely.
WARNING!
Risk of infection
Hygienically reprocess the scale regularly as described in the
respective section in this document.
Make sure that the patient has no infectious diseases.
Make sure that the patient has no open wounds or infectious skin
alterations, which may come into contact with the device.
Preventing device damage NOTICE!
Damage to device
Ensure that no liquids enter the device. They can damage the
electronics.
Switch off the device before disconnecting the power supply unit
from the mains socket.
Disconnect the power supply unit from the mains socket if you
intend to not use the device for a longer period of time. Only this way
it can be ensured that the device is currentless.
Make sure not to drop the device.
Do not expose the device to any impacts or vibrations.
Perform function controls regularly as described in the relevant
section in this document. Do not operate the device if it is damaged
or not working properly.
Ensure that there is no heat source in the immediate vicinity. Do not
expose to direct sunlight. The excessive temperature could damage
the electronics.
Avoid rapid temperature fluctuations. When the device is
transported so that a temperature difference of more than 20 °C
occurs, it must stay turned off for at least 2 hours before it can be
turned on again. Otherwise, condensation water will form which can
damage the electronics.
Use only chlorine and alcohol-free disinfectants which are explicitly
suitable for acrylic sheet and other sensitive surfaces (active
ingredient: quaternary ammonium compounds, for example).
Do not use aggressive or abrasive cleaning agents.
Do not use organic solvents (e.g. white spirit or petroleum spirit).