9.4 CHECKING THE WALL STAND
1. Disassemble Front covers of Wall Stand and clean steel cable and
brakes.
2. Check anchoring screws, they should be strongly tighten.
3. Check the vertical lock.
4. Loose the carriage lock and verify that vertical motion is released and
counterweighted at any point.
5. Take out the Cassette film tray and check correct functioning.
9.5 CHECK OFF MAINTENANCE REVISION TABLE
Hospital name: Date:
System Name: System Number ID:
Customer: Field Engineer Name:
Floor Mounted Tube Stand
Maintenance activity Comments
Electrical cables and connections
-
Mechanical cables
-
Brakes, locks and detents
-
Functional Check
-
Adjustment and calibration
-
Replacement
-
Lubrication
-
Cleaning
-