EasyManua.ls Logo

Seko KRONOS 20 pH/ORP - pH Calibration Procedures; Automatic pH Calibration; Manual pH Calibration

Default Icon
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
0000138014 rev. 1.0 EN
9
pH Calibration Menu
You can choose the automatic or manual calibration mode, in both cases the calibration at pH 7 is automatic.
- Automatic calibration:
When the display shows the value of the buffer solution, insert the probe into the bottle, press and the
display will show the countdown of the 60 seconds required to complete the calibration. If the quality of the
alignment is less than 50%, an error appears on the display; press to exit the calibration (after 4 seconds the
pump will automatically exit the calibration); if the quality is higher than 50%, the value appears on the display;
pressing you will be prompted for the pH 4 or 9 buffer solution; at this point the procedure is similar to the
previous calibration.
- Manual calibration:
When the display shows the value of the buffer solution, insert the probe into the bottle, press and the
display will show the countdown of the 60 seconds required to complete the calibration. If the quality of the
alignment is less than 50%, an error appears on the display; press to exit the calibration (after 4 seconds the
pump will automatically exit the calibration); if the quality is higher than 50%, the value appears on the display;
pressing on the display will flash the pH 7.00 value; press to insert the value of my solution, then press
to confirm and start the calibration procedure as described previously.
PROG
Cong.
Calibr.
Probe
Calibr.
Auto.
Auto Cal
7pH
WAIT
60 s.
Error
7 pH
7 pH
100% ok
Auto Cal
4/9.25pH
WAIT
60 s.
Error
4 pH
4 pH
100% ok
Calibr.
Manual
Man Cal
7pH
WAIT
60 s.
Error
7.00 pH
7 pH
100% ok
Man Cal
7.00 pH
WAIT
60 s.
Error
4.00 pH
4 pH
100% ok

Related product manuals