Paragraph 3 – Proportional dosing for the Redox (O.R.P.) measurement
Programming
Operation
The pump measures and controls the Redox value of a solution, by
programming the following parameters: set-point, set-point type,
proportional band and alarm band.
Set-point type: maximum
0
mV
s/m
max
flow-rate
Ala rm Ba nd
Ala rm Ba nd
Pr o p o r t i o n a l
Ba n d
set-point
A
l
a
r
m
O
n
A
l
a
r
m
O
n
Pump Off
P
u
m
p
O
n
Set-point type: minimum
0
mV
s/m
max
flow-rate
Ala rm Ba nd
Ala rm Ba nd
Pr o p o r t i o n a l
Ba n d
set-point
A
l
a
r
m
O
n
A
l
a
r
m
O
n
Pump Off
P
u
m
p
O
n
Can also be programmed:
- The O.F.A. (Over Feed Alarm) time in minutes, a set period of time after
which, if the measured Redox value does not reach the set-point, an
alarm signal is activated.
- Activation / deactivation of the calibration procedure
Display during operation
Allarmi e stati
• Cal = calibrazione non
eseguita
• Liv = Allarme livello
• Fls = Allarme flusso
• Alm = Misura fuori
Banda allarme
• OFA = Allarme O.F.A.
Valore misura del
potenziale Redox.
Stato della pompa
• Vuoto = pompa in start
• Stop = pompa ferma
• Paus = pompa in pausa
Valore del dosaggio in corso
(dipende dall’unità di misura
selezionata)
• Percentuale, Frequenza,
l/h, Gph, ml/m
465mV
P 100%
PROG
Config.
Config.
Funct.
Funct.
Redox
Setpoint
0mV
Set. Type
High
ProBand
100mV
AlarmBnd
20mV
OFA
0m
Calibr.