EasyManua.ls Logo

Seko KRONOS 20 pH/ORP - Proportional Redox (O.R.P.) Dosing; Redox Proportional Dosing Configuration; Additional Redox Settings; Redox Measurement Display

Default Icon
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
0000138014 rev. 1.0 EN
5
Paragraph 3 Proportional dosing for the Redox (O.R.P.) measurement
Programming
Operation
The pump measures and controls the Redox value of a solution, by
programming the following parameters: set-point, set-point type,
proportional band and alarm band.
Set-point type: maximum
0
mV
s/m
max
flow-rate
Ala rm Ba nd
Ala rm Ba nd
Pr o p o r t i o n a l
Ba n d
set-point
A
l
a
r
m
O
n
A
l
a
r
m
O
n
Pump Off
P
u
m
p
O
n
Set-point type: minimum
0
mV
s/m
max
flow-rate
Ala rm Ba nd
Ala rm Ba nd
Pr o p o r t i o n a l
Ba n d
set-point
A
l
a
r
m
O
n
A
l
a
r
m
O
n
Pump Off
P
u
m
p
O
n
Can also be programmed:
- The O.F.A. (Over Feed Alarm) time in minutes, a set period of time after
which, if the measured Redox value does not reach the set-point, an
alarm signal is activated.
- Activation / deactivation of the calibration procedure
Display during operation
Allarmi e stati
Cal = calibrazione non
eseguita
Liv = Allarme livello
Fls = Allarme flusso
Alm = Misura fuori
Banda allarme
OFA = Allarme O.F.A.
Valore misura del
potenziale Redox.
Stato della pompa
Vuoto = pompa in start
Stop = pompa ferma
Paus = pompa in pausa
Valore del dosaggio in corso
(dipende dall’unità di misura
selezionata)
Percentuale, Frequenza,
l/h, Gph, ml/m
465mV
P 100%
PROG
Cong.
Cong.
Funct.
Funct.
Redox
Setpoint
0mV
Set. Type
High
ProBand
100mV
AlarmBnd
20mV
OFA
0m
Calibr.
Probe

Related product manuals