If Dark Calibration Position Conrmation screen appears, set the Light Selection
Ring
❷
to Dark Calibration position and perform calibration again.
*
If "Please set Light Selection Ring for dark
calibration" is displayed, set the Light
Selection Ring
❷
to the dark calibration
position to start calibration.
*
When taking measurements, set the Light
Selection Ring
❷
to the range L (icon at
the middle) for ALL ambient light.
For flash light, set the ring to "H" (icon
at the top) high-power (brighter than
640lx·s) ash or "L" (icon at the middle)
for low-power ash (lower than 640lx·s).
Dark Calibration Position
Confirmation Screen
NOTE
Dark calibration is performed when new batteries are used, 24 hours passed
since last use or there is a big change in temperature between turning power
OFF and ON.
Except the cases above, dark calibration after power ON is skipped.