Battery Replacement
The Ampstik+ is powered by a single 9V battery. The battery should be re placed
when the “LO BAT” indication shows on the display. To change the battery, loosen
the screw on the battery cover at the rear of the unit. Pull the battery out of the
compartment and install a fresh battery. Secure the cover by closing and tightening
the screw on the battery cover.
Cleaning
The Ampstik+ can be cleaned by wiping with a silicone hot stick wipe to remove
sand, salt, and dirt.
High Voltage Operation
This instrument is designed to operate in high voltage fi elds. How ev er, diffi culty may
be experienced when excessive corona to the instrument occurs, resulting in a lost
reading when in HOLD mode or require the power to be cycled.