EasyManua.ls Logo

Sentec tcPCO2 - V-Check Mode Operation

Sentec tcPCO2
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 35
.
Patient Monitoring with the SDMS
If the sensor is stabilized on the skin, the RHP-reference value
can be set either a) by using the respective function in the
‘Quick Access Menu’ that activates after pressing the Enter
Button when a measurement screen is active or b) by chang-
ing the menu-parameter ‘Heating Power Mode’ from ‘Relative’
to ‘Absolute’ or ‘OFF’ and back to ‘Relative’.
To clear/reset the RHP-reference value, either remove the sen-
sor from the patient and insert it into the Docking Station or
set the menu-parameter ‘Relative Heating Mode’ to ‘OFF’.
‘V-Check™ Mode’
In standard configuration, the SDM’s numeric values and
online trends provide continuous monitoring of the enabled
parameters. If the menu-parameter ‘V-Check™ Mode’ is set
to ON (only selectable if enabled by the institution), the SDM
provides a Ventilation Spot Check with a statistical result
screen displaying mean, minimum, maximum, median and
standard deviation for the enabled parameters.
A V-Check™ Measurement consists of the V-Check™ Stabili-
zation Phase (default duration 8 minutes) and the V-Check™
Measurement Phase (default duration 2 minutes). If the
V-Check™ Measurement is finished two short signal tones
sound and the V-Check™ Results Screen activates, displaying
the above mentioned statistical results for the data assessed
during the V-Check™ Measurement Phase. The V-Check™ Re-
sults Screen remains displayed until the Menu or Display But-
ton are pressed or another V-Check™ Measurement is started.
Note: The ‘V-Check™ Mode Indicator’ appears on the ‘Ready
for use’ and ‘Calibration’ screen (p. 22) if the V-Check™
Mode is ON. On measurement screens (p. 31), the V-Check™
Down-Counter (format hh:mm:ss) is displayed on the very right
of the Status Bar (p. 48). This down-counter indicates the
duration of the V-Check™ Measurement if the V-Check™ Mea-
surement has not yet been started, the remaining time to fin-
ish the V-Check™ Measurement during an ongoing V-Check™
Measurement, and 00:00:00 once the V-Check™ Measurement
is finished. If the SDMS is not ready for use, it indicates --:--:--.
Note: Print-out of the trend curves (including the statistical
results) is automatically activated upon completion the
V-Check™ Measurement if the protocol ‘Serial Printer’ is
selected and a printer is connected to the SDM.
Note: The SDM automatically stores V-Check™ Events in its
internal memory at the start and at the end of each V-Check™
Measurement Phase. After trend data download to V-STATS™
the start and end of a V-Check™ Measurement Phase are
visualized by two colored triangles and it is possible to generate
a report which includes the same information as is provided on
the SDM’s V-Check™ Results Screen.

Related product manuals