Sentry Sample Sentry II 17
Flow Rates
Figure 10. Flow Rates
Flow Ra tes
Blowdown
1500 cc/min
Figure 10 shows the ow rates setup screen. The rst set point displayed is the blowdown
ow rate. To adjust the set point, use the ▲ and ▼ keys. To change the set point in counts
of 100, hold the MENU key while using the ▲ and ▼ keys. When nished changing the
displayed set point, press the ENTER key to display the next set point. After setting the last
ow rate, pressing ENTER returns the unit to the setup screen. The following ow rates can
be programmed:
Blowdown: This ow rate is used during the Blowdown state. During this time, the entire
sample ows to the grab sample, preventing ow to analyzers. This mode is designed to
ush the sample at a slightly higher ow rate than the sample state.
 NOTE
EPRI recommends a blowdown velocity no greater than 125% the normal sampling velocity.
Sample: This ow rate is used during the Sample state. During this time, sample is allowed
to ow to analyzers.
Shutdown: This ow rate setting is used to determine when the system should shut
down. When the total ow is less than this set point and the PRV is full open, the system
shuts down.
 NOTE
If the desire is for the system to never shut down, set the Shutdown ow rate to 0 cc/min, and
disable the Shutdown pressure setpoint.
Blowdown Time
Figure 11. Blowdown Time
Blowdown Time
hh :mm: ss
0:45 :00
Figure 11 shows the blowdown time setup screen. This setting controls the duration the
system stays in blowdown after startup. To change the seconds eld, use the ▲ and ▼
keys. To change the minutes eld, hold the ENTER key while using the ▲ and ▼ keys. To
change the hours eld, hold the MENU key while using the ▲ and ▼ keys. To return to the
setup screen, press the ESC key.
 NOTE
To disable blowdown, set the blowdown time to 0:00:00.