Page #6 background image
6 Sentry Equipment Corp
General Description
Flow control of water and steam sampling is important to help assure the quality of
the samples and repeatability of the analytical results. Constant ow in the sample line
minimizes the chance of modifying the chemistry of the samples of high purity water
through entrapment and/or release of soluble species from deposits in the system.
Sample inlet temperature and ow are easily controlled under constant conditions.
However, since sample pressures typically uctuate, maintaining a representative sample
can be dicult and labor intensive. The traditional manual sample conditioning system
can require constant adjustment during startup or as plant conditions change.
Sample Sentry II is designed to meet requirements for representative sampling by
maintaining a constant ow rate with varying inlet conditions. Automated ow control is
achieved by continuously monitoring the sample ow rate and controlling a motorized
pressure reduction valve (PRV). An actuated ball valve and solenoid valve controls the
ow path, allowing for automatic startup and shutdown of the system. Flow control is also
available in a manual mode, allowing the operator to control the system with a keypad
on the controller. Sensors allow the controller to monitor temperature, pressure, and ow
and shutdown the system when an alarm occurs. Programmable options in the controller
allow Sample Sentry II to conform to a variety of sampling applications.