EasyManuals Logo

Sharp 21ES251E User Manual

Sharp 21ES251E
25 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #13 background imageLoading...
Page #13 background image
13
13-1 13-2
PAL CHROMA ADJUSTMENT
NO
ADJUSTMENT POINT
ADJUSTMENT CONDITION / PROCEDURE WAVEFORM OR OTHERS
1 SUB COLOUR (1) Receive the "PAL Color Bar" signal. Cy G B
(I2C BUS CONTROL) (2) Press R/C to set Picture Normal condition.
(to be done after
sub-picture, sub-tint adj)
(3) Connect the oscilloscope to
R-Amp Transistor Base
(JUMPER 401)
Range :
100mV/Div
(AC) (Using 10:1 Probe)
Sweep Time :
10µsec/Div
75% W Y 100% W Mg R
(4) Using the R/C call
V05
in SERVICE mode. Adjust
V05
bus data, so that the 75% White & Red portions of PAL Color Bar be
at the
same level
shown as Fig 1-1. Fig. 1-1
(5) Clear the SERVICE mode.
NTSC CHROMA ADJUSTMENT
NO
ADJUSTMENT POINT
ADJUSTMENT CONDITION / PROCEDURE WAVEFORM OR OTHERS
1 SUB-TINT (1) Receive the "NTSC 3.58 Color Bar" signal thru AV in.
straight line slope
(I2C BUS CONTROL)
(2) Connect the oscilloscope to B-AMP Transistor Base
(JUMPER 410)
.
Range :
100mV/Div
(AC)
Sweep time :
10µsec/Div
(3) In Service mode, go to
V07, then
press
R/C Y-mute (Hex E4)
or
FLASHBACK
key.
(4) Call the "
V07
" data in service mode. Adjust the "
V07
" bus
data to obtain the waveform shown as Figure 1-1.
(5) Disable
Y-Mute
by pressing key
(Hex E4
) or
FLASHBACK
, then clear the
SERVICE mode. *NOTE : Please make sure waveform is adjusted until getting
good step slope which is can be a straight line slope at the
top of waveform step.
W Y Cy G Mg R B
Fig.1-1
SECAM CHROMA ADJUSTMENT
NO
ADJUSTMENT POINT
ADJUSTMENT CONDITION / PROCEDURE WAVEFORM OR OTHERS
1 SECAM 1) Receive "SECAM ALL WHITE" signal.
BLACK LEVEL
R-Y / B-Y
2) In the service mode, select service data
V14
.
Fig 2(a)
(I2C BUS CONTROL)
3) Connect oscilloscope to
TP 801
.
Offset
Range:
20mV/Div
(AC) (use 10:1 probe)
Sweep time:
20usec/Div
4) Adjust the
V14
so that the offset of R-Y to minimum, shown in
Fig 2(b)
, it means
adjust the offset of between No signal line and Signal line to minimum
Fig 2(b)
Offset
5) In the service mode, select service data
V15
.
6) Connect oscilloscope to
TP 801
.
Range:
20mV/Div
(AC) (use 10:1 probe)
Sweep time:
20usec/Div
7) Adjust the
V15
so that the offset of B-Y to minimum, shown in
Fig 2(b)
, it means
adjust the offset of between No signal line and Signal line to minimum
become 1 line

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sharp 21ES251E and is the answer not in the manual?

Sharp 21ES251E Specifications

General IconGeneral
BrandSharp
Model21ES251E
CategoryTV
LanguageEnglish

Related product manuals