Chapter 11 Troubleshooting
11-6
Dataletty system does not enter ready mode.
Confirm whether the Dataletty unit is online.
An error other than those above occurred.
Switch the AGS-X power OFF, then switch it ON again after about 10 seconds.
If the error continues after repeated attempts, it could be a hardware failure.
Contact your Shimadzu service representative.
In addition to system errors, if an error code other than indicated above is displayed, there
may be a hardware failure. Contact your Shimadzu service representative.
11.3 Other Problems
This section describes examples of problems that do not result in an alarm message.
1. Power Supply Problems
Cannot switch the instrument power ON.
Confirm whether a circuit breaker has tripped on the power supply panel.
Confirm whether the power cord is disconnected.
Confirm whether the input voltage mode switch is set properly.
Confirm whether a fuse has failed.
If replacing fuses, contact your Shimadzu service representative.
Confirm whether the input voltage mode switch is set properly.
Confirm whether the fuses current rating is correct.
(100 V/120 V → 10 A, 220 V/240 V → 5 A)
Confirm whether a time-delayed type of fuse is being used. (A "fast-acting" fuse
cannot be used.)
If replacing fuses, contact your Shimadzu service representative.
A ground fault circuit interrupter was tripped on the power supply panel.
There may be an electrical leak within the AGS-X unit. Immediately stop using
the system and contact your Shimadzu service representative.