M054-E093J
Instruction Manual
for
Electronic Balance
AW series AW320, AW220, AW120
AX series AX200, AX120
AY series AY220, AY120, AY62

Table of Contents

Questions and Answers:

Shimadzu AW220 Specifications

General IconGeneral
BrandShimadzu
ModelAW220
CategoryScales
LanguageEnglish

Summary

Safety Precautions

Critical Warnings

Critical safety warnings indicating potential severe injury or death if not heeded.

Operational Cautions

Safety cautions for potential minor injury or equipment damage.

Key Functionality

Balance Installation Guide

Power Supply Requirements

Specifies the required power supply voltage and adapter for the balance.

Optimal Installation Site Selection

Guidelines for selecting an optimal and safe location for balance installation.

Step-by-Step Installation Procedure

Step-by-step instructions for setting up the electronic balance.

Initial Operation Procedures

Warm-Up Procedure

Procedure for allowing the balance to stabilize before use.

Important Operational Precautions

Important notes and guidelines to observe for safe and correct operation.

Measurement Procedures

Setting the Measuring Mode

Configuring the balance to the appropriate measurement mode.

Performing a Measurement

Detailed steps for performing a weighing measurement.

Menu Navigation

Function Selection Menu

System Settings Menu

Standard Span Calibration Method Selection

Procedure to choose the type of span calibration to be performed.

Interface Settings Menu

Unit Selection Menu

Solid Specific Gravity Measurement

Liquid Specific Gravity Measurement

Add Load Weighing Function

Standard Span Calibration

Setting the Calibration Type

Procedure to set the default calibration type for the balance.

Executing Span Calibration

Steps to perform the standard span calibration.

External Weight Value for Span Calibration

External Weight Value for Built-in Calibration

Built-in Weight Calibration

Procedure for Internal Weight Calibration

Step-by-step guide for calibrating the balance using its internal weight.

GLP/GMP Output Settings

Automatic Span Calibration

Scheduled Calibration (Clock-CAL)

Enabling Scheduled Calibration

Configuration for automatic span calibration based on internal clock times.

Setting Scheduled Calibration Times

Defining Clock-CAL Start Times

Procedure to set the specific times for automatic span calibration.

Solvent Specific Gravity Setup

Sinker Volume Setup for Liquid Gravity

Performance Verification

Measurement Accuracy Tests

Repeatability Testing Procedure

Procedure for testing the consistency of measurements by repeated weighing.

Corner Load Error Testing

Procedure to test the balance's accuracy when loads are off-center.

Balance Maintenance

Troubleshooting Common Issues

Diagnosing and Resolving Operational Problems

Guidance for diagnosing and resolving common operational issues.

Error Message Reference

Error Codes and Corrective Actions

List of error messages with descriptions and corrective actions.

Technical Specifications

Balance Model Specifications

Technical details and performance characteristics of the balance models.

Windows Direct Data Transfer

Troubleshooting WindowsDirect Communication

Resolving WindowsDirect Communication Issues

Specific troubleshooting steps for issues with the WindowsDirect communication feature.

Data Communication Formats

Command Code Reference

Peripheral Device Command List

Reference list of commands for controlling the balance via peripheral devices.

Initial Balance Setup

Balance Calibration Procedure

Performing Span Calibration

Detailed steps for performing span calibration on the balance.

Menu Navigation Map