ADM-880C 07/20/0912
CONFIGURE
This message is flashed if the user presses ALT / MODE / ALT / MODE and indicates that the configuration loop has
been selected by the user.
CORR
Part of the FLO-HOOD / CORR message sequence. This message sequence indicates that the meter is taking a
FlowHood reading that is compensated for backpressure..
DATA ERROR
This message indicates that a number has been entered which is not appropriate for that type of data entry, such
as a three for the first digit of the hour setting for time.
DATE s
This is the fourth option available when the user presses the MODE key repeatedly after selecting the CONFIGURE
mode, and is used to change the date settings in the meter.
DENSITY / CHANGE / NOT ALLOWD
This message is displayed if the density selection is changed while readings are being stored in an air flow or velocity
measurement mode. A change in the density selection while storing air flow or velocity readings requires the user
to select a new memory group, as readings stored with local and standard densities must not be averaged together.
DENSITY s
This is the second option available when the user presses the MODE key repeatedly after selecting the CONFIGURE
mode and is used to change the density settings in the meter.
DIFF PRES
This signal will be flashed when the differential pressure measurement mode is selected, and upon each subsequent
operation of the READ key. It is also displayed as part of the sequence of displays when the associated readings
are being viewed by pressing ALT / VIEW following a pitot tube velocity reading. It precedes the display of the
velocity pressure read by the meter to produce the pitot tube reading.
D PRESS IN
This message is displayed as part of the sequence of displays when associated readings are being viewed following
a differential pressure reading.
ENGLISH
This message indicates that the readings will be in English units. The meter will save the units selection (metric or
English) and will automatically default to the selected units the next time it is turned on.
ENGLISH s
This prompt indicates that English units are being selected using the meter configuration loop.
ERASE s
This message is displayed when the operator has pressed ALT / STORE / ALT / STORE, followed by pressing the
STORE key twice while viewing a stored reading. The reading displayed will be erased when the operator presses
the READ/SELECT key.
ERASING
This message is displayed when the selected reading displayed using a SCAN key is being erased.
ESCAPE
This message is displayed when the user presses ALT / ALT. The meter will revert to the prior selections.
FLO-HOOD
This message will be flashed when the FlowHood measurement mode is selected, or when the meter is first turned
on, if the flaps plug is connected.
FLO-HOOD / CORR
This message sequence indicates that the meter is taking a FlowHood reading that is compensated for backpressure.
FLO-HOOD / UNCORR
This message sequence indicates that the meter is taking a FlowHood reading that is not compensated for
backpressure.