5.5.1 Starting the measurement
Once the power is supplied and no abnormal issues are observed,
the meter is ready to perform the measurements. The default
display is for the mass flow measurement having two numerical
lines on the LCD. The upper line is the accumulated or totalized
flow rate, and the lower line is the instant flow rate. If the battery
symbol turns into , proceed to change the battery or plug in
the external power cord.
5.5.2 MENU entry with a verified password
At the flow measurement (main) display, press the central “M” MENU key, it will enter into the
password setting and verification MENU. The default password is “11111”. If the password is incorrect,
the display will return to the main display.
To enter a new password, press the “Up” or “Down” key to change the digit that flashes, and press
the “M” key to confirm. Repeat this process for all 5 digits and the meter will enter into the menu
setting interfaces/screen.
Subsequently, the MENU allows the user to:
Set Modbus address F2 - Addr
Change communication baud rate F3 - bPS
Reset or calibrate the offset F11 - oFFST
Enter the gas conversion factor (GCF) F12 - GCF
Change the display mode F38 - dISP
Change the display language F39 - LAnG
Open/Close the valve F43 - VALVE
Set upper flow rate limit alarm F51 - ALM-H
Set lower flow rate limit alarm F52 - ALM-L
Set totalized or accumulated flow rate alarm F53 - ALM-A
Change the default password F91 - PASS
Clear or reset the totalizer or accumulated flow rate F92 - CLr-A
Exit from the MENU F99 - qUIT
Note: During this process, the meter will continue to measure the flow without being interrupted.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siargo MF5700 Series and is the answer not in the manual?

Siargo MF5700 Series Specifications

General IconGeneral
BrandSiargo
ModelMF5700 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals