Serial number of the product
Gas correction factor (R/W)
Accumulated/totalized flow rate alarm
Upper flow rate limit alarm
Reset accumulated or totalized flow rate (W)
Write protection of selected parameters (W)
The detailed information of each register is described below: Y: enabled; N: disabled
Flowmeter address 0x0081
Values from 1 to 255 except for 157 (0x9d). 0 is the broadcast address.
SN, Serial number 0x0030
Series Number of the product, SN
Notes
SN= value(0x0007), value(0x0008),….,value (0x000C);
Receiving 12 bits as: 2A 47 37 41 45 49 30 32 30 35 38 2A , the corresponding
Serial Number is *G7AEI02058*.
Current flow rate 0x003A ~ 0x003B
Notes
e.g.: for a flow rate of 20.340 SLPM, the user will read “0” from register
0x003A and “20340” from register 0x003B, therefore
Current flow rate = (0*65536+20340)/1000 = 20.340