EasyManuals Logo

Siemens ARCADIS Avantic Maintenance Instruction

Siemens ARCADIS Avantic
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #13 background imageLoading...
Page #13 background image
4 - 1
Siemens AG SPR2-330.662.01 Page 1 of 2 ARCADIS Avantic
Medical Solutions Rev. 02 05.06 CS PS SP
Checking the function blocks and displays 4
1 Checking the ”deadman” switch function 4
1.1 Exposure 4
An exposure value is set at the control console.
- Activate the exposure with the exposure switch and end it immediately by releasing the
switch.
The exposure must end immediately.
1.2 Fluoroscopy 4
Activate fluoroscopy for a short time and release the fluoroscopy switch.
Fluoroscopy must stop immediately.
a) with the fluoroscopic switch at the control console
b) with the foot switch
2 Radiation indicators 4
Check that the radiation indicators at the control console and on the monitor cart are visi-
ble during radiation and that an audible (whistling) signal is sounded after exposure
release for about 1.5 seconds.
Deadman’s switch is functioning YES NO
Radiation indicators are functioning YES NO

Other manuals for Siemens ARCADIS Avantic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Avantic and is the answer not in the manual?

Siemens ARCADIS Avantic Specifications

General IconGeneral
BrandSiemens
ModelARCADIS Avantic
CategoryMedical Equipment
LanguageEnglish

Related product manuals