20 Function check and final work steps
ARCADIS Varic/Orbic SPR2-310.812.03.01.02 Siemens AG
09.06 CS PS SP
Page 20 of 20
Medical Solutions
Final work steps 3.1
Covers 0
Close all cover panels if any are still open. Make sure the protective conductors make
good contact.
Protective conductor resistance test 0
Observe the instructions for the protective conductor resistance test in Chapter 1 of
these instructions.
Perform the protective conductor resistance test according to ARTD-002.731.17...
¹ The protective conductor resistance must not exceed 0.2 ohms.
Customer training 0
NOTE
The system operating manual already describes the multi-func-
tion footswitch and its functions.
Explain the functions of the multi-function footswitch to the customer. Also, demon-
strate how to place the footswitch in the holder.