Siemens AG SPR2-310.812.03.01.02 ARCADIS Varic/Orbic
09.06 CS PS SP
General information and safety guidelines 5
Page 5 of 20
Medical Solutions
Safety information 1.1
General safety information 0
WARNING
Danger of injuries, death or material damage.
Non-compliance can lead to death, injury or material damage.
Please note:
¹ The product-specific safety notes in these instructions,
¹ The general safety information in TD00-000.860.01... and
¹ The safety information in accordance with ARTD Part 2.
General electrical safety information 0
WARNING
Electrical safety!
Non-compliance can lead to severe injury or even death, as well
as material damage.
¹ Parts under electrical voltage are accessible when the
covers are open. To avoid danger, disconnect the system
from the power supply before opening the covers. Dis-
connect the power plug.
¹ If an uninterruptible power supply (UPS) is installed in
the system, the voltage output of the UPS must also be
deenergized or the voltage output plug must be discon-
nected.
¹ If work steps must be performed using electrical power,
the general safety information according to
TD00-000.860.01 must be observed.