ID/
Icons
Diagnostic Action Comment
128 Standard density Value above alarm limit. Check process condi‐
tions. Align Upper alarm limit to normal process
conditions.
129 Standard density Value above warning limit. Check process con‐
ditions. Align Upper warning limit to normal
process conditions
130 Standard density Value below warning limit. Check process con‐
ditions. Align Lower warning limit to normal
process conditions.
131 Standard density Value below alarm limit. Check process condi‐
tions. Align Lower alarm limit to normal process
conditions.
132 Standard volume flow Value above alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter 'Upper alarm limit'
Only hydrocarbon
133 Standard volume flow Value above warning limit. Check process con‐
ditions or align limit to normal operation. Adjust
parameter 'Upper warning limit'
Only hydrocarbon
134 Standard volume flow Value below warning limit. Check process con‐
ditions or align limit to normal operation. Adjust
parameter 'Lower warning limit'
Only hydrocarbon
135 Standard volume flow Value below alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter 'Lower alarm limit'
Only hydrocarbon
136 Totalizer 1 Value above alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter 'Upper alarm limit'
Diagnosing and troubleshooting
10.2 Fault codes and corrective actions
MASS 2100 & FC300 (FCT030)
198 Operating Instructions, 06/2017, A5E39789040-AA