Siemens AG SPR8-230.840.01.05.02 MOBILETT XP
11.05 CS SD 24
Service programs 71
Page 71 of 72
Medical Solutions
Program 11 - Exposure points 0
Program 12 - system type 0
Program 13 - sound level of the keyboard 0
Program 11
After the start of the service program, "ON" or “OFF“ is shown on the display.
ON = 49 (for ECO = 47) exposure points possible (default factory setting); OFF = 25
(for ECO = 24) exposure points possible
Use the mAs+ or mAs- keys to switch over.
Program 12
After the start of the service program, system information is shown on the display.
E.g.:
- “DIG 102” = Digital, 102 = Firmware VA00B (A = 1, B = 2, ...)
- “103” = XP Hybrid, 103 = Firmware VA00C
- “ECO 102” = ECO, 102 = Firmware VA00B
Program 13
After the start of the service program, "bep1", “bep2”, or “bep3“ is shown on the dis-
play.
bep1 = quiet; confirmation tone upon pressing the "kV," "mAs," or "collimator lamp"
keys is OFF (factory default setting).
bep2 = quiet; confirmation tone upon pressing the "kV," "mAs," or "collimator lamp"
keys is ON
bep3 = loud; confirmation tone upon pressing the "kV," "mAs," or "collimator lamp"
keys is ON