14 System-related Adjustments
Display TD00-000.841.02.04.02 Siemens AG
05.05 CSPS21
Page 14 of 22
Medical Solutions
SP Application 0
Mammomat Novation DR 0
Replacing the TFT Monitor
It must be noted that the cabling on the back of the unit must be made in accordance with
the designations.
Selecting the Test Images
The test images are located under "Service\Acquisition System\Quality Tests" with the
designation "Display". Start the service software for this purpose.
Select Service and Local Service in the menu bar under Options.
Delete the password in the SW screen and click on OK.
Fig. 8: Mammomat Novation DR -1 Service SW