6 General Information
Display TD00-000.841.02.04.02 Siemens AG
05.05 CSPS21
Page 6 of 22
Medical Solutions
Adjustment Note 0
Tab. 1 Push buttons
NOTE
Adjustments may be performed only after a warm-up time of 15
minutes!
NOTE
As a rule, the push-buttons are not locked in replacement units
(from the E-warehouse).
They must be locked following installation.
All required adjustments are made using the push-buttons on
the side of the unit.
The push-buttons are locked when shipped (new system).
For the function of the push-buttons, see the Instruction Man-
ual included in the shipment.
Unlocking or locking them is done as follows:
- Press the “SET” button 1x.
- Press the “ Up” button (arrow pointing up) 3x.
Fig. 1: