Siemens AG TD00-000.841.02.04.02 Display
05.05 CSPS21
System-related Adjustments 7
Page 7 of 22
Medical Solutions
3-
2System-related Adjustments
General Remarks regarding System-related Adjustments 0
NOTE
• For the function of the push-buttons, see the Instruction Manu-
al included in the shipment.
• All values listed, including the listed tolerances, are "setting
values”.
The listed tolerances are not values for the constancy check of
the display!
• The monitor must be switched on at least 20 minutes prior to
beginning the adjustment.