Siemens AG TD00-000.841.02.04.02 Display
05.05 CSPS21
System-related Adjustments 19
Page 19 of 22
Medical Solutions
The AWS test images and TFT Auto Calibration can now be used to adjust the frequen-
cy and phase.
Fig. 13: Mammomat Novation DR - 6 Service SW
Pos. 1 AWS Test Images
Pos. 2 TFT Auto Calibration
Fig. 14: Vertical bars on SMPTE
Pos. 1 Max. resolution