SIREMOBIL Compact Register 3 RXR2-130.033.04 Page 2 of 4 Siemens AG
Rev. 01 08.96 TD SP 1 Medical Engineering
2
Installing and setting the Video Recorder1
General 1
Safety information 1
Checks and adjustments which must be performed with radiation switched ON are identi-
fied by the radiation warning symbol . During these types of adjustments, radiation pro-
tective clothing must be worn.
Protective conductor test 1
Perform the protective conductor test according to TI 236 / ARTD - 002.731.17... after
completing all work and reinstalling all covers.
The protective conductor resistance must not exceed 0.2 ohms.
Additional Information 1
These instructions apply to retrofitting a video recorder on a SIREMOBIL Compact system
only.
Materials required 1
Measurement and auxiliary devices required 1
Tools required 1
• Tool kit (see ARTD / part 3)
• 1 set of Allen keys
When performing service work and tests, adhere to the product-specific
safety information contained in the documents as well as the general
safety information contained in Register 2 of the TI folder.
• Video recorder connection kit 31 00 489 G5437
• Service PC
• Service software (see Log book)
• Connecting cable (from service PC to host) 99 00 440 RE999
• Protective conductor meter 44 15 899 RV 090