SIREMOBIL Compact Register 3 RXR2-130.033.04 Page 2 of 4 Siemens AG
Rev. 01 08.96 TD SP 1 Medical Engineering
2
Installing and setting the Video Recorder1
General 1
Safety information 1
Checks and adjustments which must be performed with radiation switched ON are identi-
fied by the radiation warning symbol . During these types of adjustments, radiation pro-
tective clothing must be worn.
Protective conductor test 1
Perform the protective conductor test according to TI 236 / ARTD - 002.731.17... after
completing all work and reinstalling all covers.
The protective conductor resistance must not exceed 0.2 ohms.
Additional Information 1
These instructions apply to retrofitting a video recorder on a SIREMOBIL Compact system
only.
Materials required 1
Measurement and auxiliary devices required 1
Tools required 1
Tool kit (see ARTD / part 3)
1 set of Allen keys
When performing service work and tests, adhere to the product-specific
safety information contained in the documents as well as the general
safety information contained in Register 2 of the TI folder.
Video recorder connection kit 31 00 489 G5437
Service PC
Service software (see Log book)
Connecting cable (from service PC to host) 99 00 440 RE999
Protective conductor meter 44 15 899 RV 090
CAUTION