Installing and setting the Video Recorder 3
Siemens AG Register 3 RXR2-130.033.04 Page 3 of 4 SIREMOBIL Compact
Medical Engineering Rev. 01 08.96 TD SP 1
Line voltage cable 1
• Connect the video recorder line voltage cable at terminals X2.1 and X2.2.
• Connect the protective conductor to the protective conductor terminal strip (ge/gn).
• Route the video recorder line voltage cable to the video recorder compartment.
Video cable 1
With an existing multiformat camera or video printer 1
• Remove the 75 Ohm termination from the video output of the muliformat camera or
switch off the 75 Ohm termination for the video printer.
• Connect the video cable at the video output of the multiformat camera or the video
printer and route the cable to the video recorder compartment.
• Connect the video recorder. Ensure proper termination of the video cable
(75 Ohm).
Without a multiformat camera or video printer 1
• Connect the video cable to Memoskop connector "Out3".
• Route the video cable to the video recorder compartment.
• Connect the video recorder. Ensure that the video cable is terminated correctly
(75 Ohm).
Programming 1
On Memoskop C (with keyboard) 1
• Switch the system on.
• Simultaneously press keys [CTRL] and [U] on the memory keyboard.
➪ This calls the User Setup menu for Memoskop.
• Call up the "B output for video recorder" submenu.
• Perform the configuration for monitor A .
(Perform the configuration for monitor B only by special customer request).
• End the user menu.
• Switch the SIREMOBIL Compact off and on again to activate the new configuration.
Disconnect the SIREMOBIL Compact line voltage cable prior to
beginning work.