Installing and setting the Video Recorder 3
Siemens AG Register 3 RXR2-130.033.04 Page 3 of 4 SIREMOBIL Compact
Medical Engineering Rev. 01 08.96 TD SP 1
Line voltage cable 1
Connect the video recorder line voltage cable at terminals X2.1 and X2.2.
Connect the protective conductor to the protective conductor terminal strip (ge/gn).
Route the video recorder line voltage cable to the video recorder compartment.
Video cable 1
With an existing multiformat camera or video printer 1
Remove the 75 Ohm termination from the video output of the muliformat camera or
switch off the 75 Ohm termination for the video printer.
Connect the video cable at the video output of the multiformat camera or the video
printer and route the cable to the video recorder compartment.
Connect the video recorder. Ensure proper termination of the video cable
(75 Ohm).
Without a multiformat camera or video printer 1
Connect the video cable to Memoskop connector "Out3".
Route the video cable to the video recorder compartment.
Connect the video recorder. Ensure that the video cable is terminated correctly
(75 Ohm).
Programming 1
On Memoskop C (with keyboard) 1
Switch the system on.
Simultaneously press keys [CTRL] and [U] on the memory keyboard.
This calls the User Setup menu for Memoskop.
Call up the "B output for video recorder" submenu.
Perform the configuration for monitor A .
(Perform the configuration for monitor B only by special customer request).
End the user menu.
Switch the SIREMOBIL Compact off and on again to activate the new configuration.
Disconnect the SIREMOBIL Compact line voltage cable prior to
beginning work.