SIREMOBIL Iso-C
System
Troubleshooting Guide
SP
2008
SPR2-230.840.01.01.02
Replaces: SPR2-230.061.01.14.02
English
Doc. Gen. Date: 08.08
n.a.CS PS SP
SIREMOBIL Iso-C
© Siemens, 2008
0282945005646950

Table of Contents

Other manuals for Siemens SIREMOBIL Iso-C

Question and Answer IconNeed help?

Do you have a question about the Siemens SIREMOBIL Iso-C and is the answer not in the manual?

Siemens SIREMOBIL Iso-C Specifications

General IconGeneral
BrandSiemens
ModelSIREMOBIL Iso-C
CategoryMedical Equipment
LanguageEnglish

Summary

Copyright Version Disclaimer

Copyright

Information about copyright ownership.

Document Version

Details on manual version and updates.

Disclaimer

Legal disclaimer regarding documentation use.

Prerequisites

Required documents

Lists necessary documentation for the procedure.

Required tools, measurement and auxiliary devices

Lists essential tools and equipment for service.

Information regarding this document

Guidance on using the service instructions.

Position designations

Explains directional terms used in the manual.

Notes and symbols

Explains meaning of symbols and emphasized text.

Safety information

Crucial safety warnings and precautions.

Danger of infection with pathogens!

Warnings about potential biological hazards.

X-ray radiation!

Safety instructions regarding X-ray exposure.

Risk of injury on mechanical parts!

Warnings about hazards from mechanical components.

Electrostatic voltage!

Precautions against ESD damage.

Danger of injury when dismantlingassembling mechanical parts!

Warnings about injury during disassembly/assembly.

Notes on the protective conductor resistance test

Guidance on performing protective conductor tests.

Notes on device leakage current measurement

Instructions for measuring device leakage current.

Covers

Procedures for removing/installing covers.

SIREPHOS cover

Steps for removing and installing the SIREPHOS cover.

Installingremoving the C-arm covers

Procedures for C-arm cover removal and installation.

Monitor trolley

Instructions for removing and installing monitor trolley covers.

I.I. service position

Steps to move the I.I. into the service position.

Installing the I.I.

Procedures for installing the I.I. component.

Concluding steps

Final steps after performing work.

Line voltage adjustment

Checking the line voltage on the monitor trolley

How to check line voltage on the monitor trolley.

Checking the operating voltages

Low voltages

Procedures for checking low voltage levels.

I.I. voltages

How to check Image Intensifier voltages.

Loading the operating software

Control board D1, host

Steps for loading software onto control board D1.

Storing the error log as an ASCII file

How to save the error log in ASCII format.

Saving trace files

Procedures for saving trace files.

Saving current trace files directly after display of an error

Saving trace files after an error occurs.

Saving the Event-Log files

Steps to save Event-Log files.

Adjustments Programs

Generator, kV offset adjustment

How to adjust the generator's kV offset.

Generator adjustment (filament circuit learning program)

Performing generator adjustment via learning program.

Checking the generator adjustment

Verifying the generator adjustment settings.

Dose rate adjustment

Procedure for adjusting the dose rate.

Reduction of maximum air kerma rate

Procedures to reduce maximum air kerma rate.

Collimator, X-Iris

Adjusting the X-Iris collimator.

Collimator, slot diaphragm

Adjusting the slot diaphragm.

Camera rotation

Procedures for adjusting camera rotation.

Dose-area product measurement device PTW

Calibration and programming for PTW dose-area device.

Dose-area product measurement device WELLHFER

Programming and testing WELLHÖFER dose-area device.

Checking the accuracy of the dose-area product measurement device

Verifying accuracy of dose-area product measurement.

Overview of SIREMOBIL Iso-C Parameters

Overview of available SIREMOBIL Iso-C parameters.

Configuring the MEMOSKOP

Configuration steps for MEMOSKOP.

MEMOSKOP, configuration and country-specific language (User Setup)

Configuring language and user setup for MEMOSKOP.

Replacing boards and components

D1 control board

Steps for replacing the D1 control board.

Power board D2

Procedures for replacing the D2 power board.

Interface board D30

Steps for replacing the D30 interface board.

Replacing the VIDEOMED DC

Procedures for replacing the VIDEOMED DC.

Basic system operating panel

Procedures for replacing the basic system operating panel.

Frequency converter FU (vertical lift)

Steps for replacing the frequency converter.

Power Supply M14, +5 V +15 V -15 V

Replacing the M14 power supply.

Power Supply M13, +13 V VIDEOMED DC

Replacing the M13 power supply.

I.I. mini-voltage supply

Procedures for replacing the I.I. mini-voltage supply.

Adjusting the I.I. mini-voltage supply

How to adjust the I.I. mini-voltage supply.

Collimator

Procedures for replacing the collimator.

SIREPHOS

Procedures for replacing the SIREPHOS.

Replacing the I.I.

Steps for replacing the Image Intensifier.

Replacing the Compact optics

Procedures for replacing Compact optics.

MEMOSKOP C-E

Configuration of MEMOSKOP C-E.

MEMOSKOP C-E 100

Configuration of MEMOSKOP C-E 100.

MEMOSKOP C MEMOSKOP C & MOD

Configuration of MEMOSKOP C/C & MOD.

MEMOSKOP C-SUB MEMOSKOP C-SUB & MOD

Configuration of MEMOSKOP C-SUB/C-SUB & MOD.

MULTISPOT 2000

Installation and configuration of MULTISPOT 2000.

Dose-area product measurement device

Procedures for replacing dose-area product devices.

Laser targeting device

Procedures for replacing the laser targeting device.

Mouse, external mouse, and PC keyboard

Procedures for replacing input devices.

Video switcher

Procedures for replacing the video switcher.

Cut-off delay - PC board D210

Replacing PC board D210 for cut-off delay.

3 D reconstruction PC

Procedures for replacing the 3D reconstruction PC.

SIMOMED HM monitor

Procedures for replacing the SIMOMED HM monitor.

3 D reconstruction system electronics

Replacing 3D reconstruction system electronics.

Orbital motor

Procedures for replacing the orbital motor.

Toothed belt of the orbital drive

Procedures for replacing the toothed belt.

Orbital drive - actual value potentiometer

Procedures for replacing the actual value potentiometer.

Angulation part

Procedures for replacing the angulation part.

Brakes Movements

Brake force brake torque

Measuring brake force and torque.

Measuring the horizontal force required for movement

Measuring horizontal force for movement.

Measuring the horizontal swivel force

Measuring horizontal swivel force.

Measuring the orbital force required for movement

Measuring orbital force for movement.

Measuring the angular force required for movement

Measuring angular force for movement.

Angulation brake magnet

Replacing the angulation brake magnet.

Horizontal swivel brake

Replacing the horizontal swivel brake.

Vertical lift brake

Procedures for replacing the vertical lift brake.

Vertical lift motor

Procedures for replacing the vertical lift motor.

Vertical lift limit switches

Replacing vertical lift limit switches.

Switch for blocking vertical lift

Replacing switch for blocking vertical lift.

Emergency off switch

Replacing the emergency off switch.

Lubricating the lifting column

Lubrication procedures for the lifting column.

Tensioning the drive belt

Procedure for tensioning the drive belt.

Supplement: measuring the tube current

Changes to previous version