© Siemens, 2008 SPR2-230.840.01.01.02 SIREMOBIL Iso-C
08.08 CS PS SP
Loading the operating software 33
Page 33 of 148
For internal use only
Storing the error log as an ASCII file 4.1
x If you store the error log using the backup function of the service program, the error log
will be stored as a binary file which cannot be read with a word editor. For easier evalua-
tion of the error log, it is also possible to store it as an ASCII file. For easier service sup-
port from the USC or HSC, please send the error log as an ASCII file.
NOTE
The parameter diskette must be in drive A before starting the
service program and during the service work. Without a
parameter diskette in drive A, the short error messages will
not be stored in the error log file and the text "no error text
available" is stored behind the error codes.
Prerequisites:
The service PC is already connected, the service program is already started.
x Select the menu "Diagnostics...."
x Select the menu "Error log".
x Click the button "Get from unit". Wait until the error log is transferred to the service PC.
x Click the button "Export to file".
x Select your hard disk drive as the target for storing the file.
x For the file name, enter the current date, the part number, and the serial number of the
unit as described in the example.
x Store the file on the hard disk drive of the service PC to a path of your choice.
x Exit the service program.
x Copy the previously stored file to a formatted, empty disk.
Example: The current date is (DD-MM-YY): 29-03-01
The unit part no. is: 28 29 450
The unit serial no. is: 01234
Enter the file name: 22-03-01_28 20 450_01234.txt