48 Adjustments / Programs
SIREMOBIL Iso-C SPR2-230.840.01.01.02 © Siemens, 2008
08.08 CS PS SP
Page 48 of 148
For internal use only
Reduction of maximum air kerma rate 5.4
WARNING
X-ray radiation!
See (Safety information / p. 11).
¹ Protect yourself against radiation exposure. Wear a
lead apron.
NOTE
Regulations in some countries require the reduction of the
maximum air kerma rate.
Depending on the country, the measuring unit for air kerma
rate is either “R/min” or “mGy/min” (10 R/min = 87 mGy/min).
This section, as an example, describes how to reduce the
maximum air kerma rate to 45 mGy/min. However, in each
country the respective limit value of that country has to be
used for calculation.
Prerequisites 0
x One of the following FL curves must be programmed: S2, HC2, or IODINE.
The FL curve S2, HC2, or IODINE must be used for the following adjustment.
x The X-ray tube cover must be installed.

Table of Contents