Adjustments / Programs 51
Page 51 of 148
For internal use only
x Select "Save to file".
¹ The programmed value has been transferred to the SIREMOBIL Compact L and
temporarily stored to the disk.
NOTE
You must exit the service program with "Logoff" and "Quit".
Only then will the modified parameters be stored on the
floppy disk as a new file.
Checking 0
X x After programming, check the maximum air kerma rate again.
¹ If the measured value still exceeds 45 mGy/min, reduce the parameter value
accordingly.