Page #105 background imageLoading...
Page #105 background image
© Siemens, 2008 SPR2-230.840.01.01.02 SIREMOBIL Iso-C
08.08 CS PS SP
Replacing boards and components 105
Page 105 of 148
For internal use only
MEMOSKOP C-E 6.13
NOTE
Check the software status of the newly delivered
MEMOSKOP. If the software status of the MEMOSKOP differs
from that of the system, the system software status must be
downloaded to the MEMOSKOP. The MEMOSKOP software
status can be read out in the 2nd line of the User Setup. The
first number describes the MEMOSKOP type. (2=CE; 3=
CE-100/ C/ C & MOD; 4=C-SUB/ C-SUB & MOD).
NOTE
Read and record the parameters for MEMOSKOP C-E if you
are able to select User Setup. After replacing the Memoskop,
configure the MEMOSKOP C-E according to the parameters
recorded.
x Replace the MEMOSKOP C-E. The attachment screws are on the underside of the
storage device.
x Check the software status of the MEMOSKOP. Download the system software to the
MEMOSKOP if necessary.
x Use the MEMOSKOP Service Software to import the language file.
x Configure the MEMOSKOP C-E for the correct video frequency and the number of
monitors present. See (MEMOSKOP C-E, keyboard emulation / p. 77), (Configuring
the video frequency / p. 77) and (MEMOSKOP, configuration for the number of existing
monitors / p. 77).
x Configure the following parameters in Technical Setup for the MEMOSKOP:
Number of Monitors Configure 1 for SIREMOBIL with 1 monitor. Configure 2 for
SIREMOBIL with 2 monitors.
Hardcopy Check whether 5000 is programmed in "4 Max Hardcopy
time". If not, program this value.
Edge 1 coef set Check whether 3 is programmed; if not, program this value.
Edge 2 coef set Check whether 8 is programmed, If not, program this value.
Motion detection setup Check whether the following are programmed - MD1:
Threshold 100 and at MD2: Threshold 120. If not, program
these values.
Date / Time format Set to country-specific format.
Keyboard Set to the keyboard layout being used.
Display of realtime clock Configure Yes. Configure No only upon customer request.
Display of rotated images Configure Yes.
Monitor zoom Configure No.
Default LUT Configure 2. Program a different LUT curve only upon cus-
tomer request.

Table of Contents

Other manuals for Siemens SIREMOBIL Iso-C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIREMOBIL Iso-C and is the answer not in the manual?

Siemens SIREMOBIL Iso-C Specifications

General IconGeneral
BrandSiemens
ModelSIREMOBIL Iso-C
CategoryMedical Equipment
LanguageEnglish

Related product manuals