106 Replacing boards and components
SIREMOBIL Iso-C SPR2-230.840.01.01.02 © Siemens, 2008
08.08 CS PS SP
Page 106 of 148
For internal use only
x Configure the following parameters in User Setup for MEMOSKOP:
x Perform the IQ quick test.
Default Edge Configure 1. Program a different edge enhancement value
only upon customer request.
Monitor split Configure Yes.
Monitor text Configure Yes.
Maximum image number n.a.
Disable dose rate selection
in user setup
Program for "No" when equipped with an dose-area product
measurement device. The menu "Storage of Dose on disk"
and the menu "Dose on Hardcopy output" are visible in
User Setup. Otherwise set to "Yes".
Unit serial number Program the serial number of the SIREMOBIL. Do not enter
the leading zero.
Product name Program "SIREMOBIL Iso-C".
Default program change-
able
Default: Yes. Depending on country-specific regulations
configure Yes or No.
Software backward com-
patible
Configure "No".
Operating mode parame-
ters
Change the parameters programmed for the organ pro-
gram upon customer request.
VCR output Configure to Monitor A.
Hardcopy Release For SIREMOBIL with 1 monitor: Configure to monitor A. For
SIREMOBIL with 2 monitors: Configure to monitor B.
Text monitor For SIREMOBIL with 1 monitor: Configure to monitor A. For
SIREMOBIL with 2 monitors: Configure to both monitors.
Split Screen Configure horizontally. Configure vertically only upon cus-
tomer request.
Display of time Configure Yes.
Set Date / Time Set the current date and time.
Hospital Name Enter the hospital name.
Storage of Dose on disk If the dose-area product measurement device is present
(Diamentor), configure Yes. Otherwise, configure No.
Dose on Hardcopy output If the dose-area product measurement device is present
(Diamentor), configure Yes. Otherwise configure No.