Adjustments / Programs 65
Page 65 of 148
For internal use only
Dose-area product measurement device "PTW" 5.11
NOTE
This adjustment is used to calibrate the dose-area product
measurement device (PTW) of the SIREMOBIL Iso-C display.
Parameter "CONSTANT (Qt * Kg)" = T
w
0
x Read and record the T
w
value shown on the electronics unit of the dose-area product
measurement device.
Programming 0
x Connect the service PC to the serial interface of the SIREMOBIL Iso-C.
x Start the service program.
x Select the Adjustment, Parameter menu.
x In the Combo box "Parameter Groups:", select "Diamentor" .
x Click "Get from Unit".
x The Parameter "1. Diamentor Existing" must be set to YES .
x Select parameter "2. Constant (Qt * Kg)".
x In "Value Actual:" field, enter the parameter T
w
.
x Click "Set Value".
x Click "Put to Unit".
x Click "Save to File".
NOTE
You must exit the service program with "Logoff" and "Quit"
to store the modified parameters to disk as a new file.