Commissioning
6.5 Local commissioning
SITRANS LT500 with ultrasonic transducer inputs
92 Compact Operating Instructions, 02/2024, A5E52021877-AA
Step: Select measurement point
Select measurement point
Sets the measurement point to be configured for alarms.
• Measurement point 1
• Measurement point 2
• External TS-3
• Exit
Step: Alarm sources
Alarm sources
Sets the source for alarms.
Select setting “Sensor diagnostics” to configure alarms based on diagnostics, such as loss of
echo (LOE).
This step will not appear if setting "External TS-3" is selected in the previous step of the
wizard.
• Level
• Space
• Distance
• Head
• Volume
• Volume flow
• Sensor temperature
• Sensor diagnostics
The list of process values available in this step is based on any configuration completed via
"Quick commissioning" wizards prior to running this wizard. For example, if a level application
is configured via the quick commissioning wizard, only the setting "Level" will be available in
this step. To set alarms for other process values, see menu "Process values" (2.2) (after
completing the alarms wizard).